MEDICAL POWER SUPPLIES
>> Dual Output |
  |
Model Number |
Output
Voltage |
Max. Output current |
Reg. |
Ripple
& Noise |
PDF |
Max. Load
(8.54 CFM) |
Max. Load
(Convection) |
X2340U-M0312F |
V1:+3.3V |
50A |
165W |
30A |
100W |
+/- 5% |
+/-1% |
|
V2:+12V |
25A |
300W |
20A |
240W |
+/- 5% |
+/-1% |
X2340U-M0324F |
V1:+3.3V |
50A |
165W |
30A |
100W |
+/- 5% |
+/-1% |
V2:+24V |
12.5A |
300W |
10A |
240W |
+/- 5% |
+/-1% |
X2390U-M0512F |
V1:+5V |
50A |
250W |
30A |
150W |
+/- 5% |
+/-1% |
V2:+12V |
25A |
300W |
20A |
240W |
+/- 5% |
+/-1% |
X2390U-M0524F |
V1:+5V |
50A |
250W |
30A |
150W |
+/- 5% |
+/-1% |
V2:+24V |
12.5A |
300W |
10A |
240W |
+/- 5% |
+/-1% |
X2400U-M1224F |
V1:+12V |
33.33A |
400W |
25A |
300W |
+/- 5% |
+/-1% |
V2:+24V |
16.66A |
400W |
12.5A |
300W |
+/- 5% |
+/-1% |
- This series are designated as X2zzzU-MyF where x can be E (enclosed type) or U (U-chassis), y can be 0312, 0324, 0512, 0524 and 1224.
- For models U chassis type which need external forced airflow min. 8.54 CFM to achieve maximum power.
- 10% minimum load is required for all outputs to maintain the ripple and regulation
- Total combined power of V1 and V2 not to exceed 400W with 10CFM forced air and 300W convection cooling for all models.
- Output is fully isolated
- Cover is optional for U chassis enclosure.
* Dimensions: B type(5"W x 9"L x 1.6"H), U type(5"W x 8"L x 1.6"H) |
| For lower or higher wattage models not listed here or if you did not find what you are looking for, please email your inquires. <Click here to send us your email inquiry> |
|